LINKMED® HL7 TCP/IP Transmitter_
Configuration

 

LINKMED® HL7 TCP/IP Transmitter (Outbound communication client) is used to transmit HL7 result or query message to HIS or Clinical systems. The HL7 TCP/IP transmitter monitors result or query folder for file, if existed automatically sends to destination..
LINKMED® HL7 TCP/IP transmitter can only be configured to send report or query message to one IP address and one port numbers. Multiple instances of the TCP/IP Transmitters are needed to send query or report to more than one IP address, call LINK Medical at 888-893-0900 or 1+ 781-453-0300 to purchase additional license(s) of the TCP/IP Transmitter
You start the Transmitter manually during initial configuration from your system server or from LINKMED® Interface server: Start>LINK Medical>LINKMED® Interface>LINK Transmitter Service

  1. TCP/IP Transmitter Configuration:
    Before you Begin
    You need the following information in order to configure the Transmitter to send HL7 result messages and receive HL7 acknowledgements
  • The IP address of your client server to which the Transmitter will be sending result or query message
  • The port number of the server to which the Transmitter can open a TCP/IP connection
  • Filename (default is *.sen) and directory folder (default C:\LINKMED\CFGLINK\ResultS\) of the system providing the HL7 result or query messages

Next>>>


LINKMED® HL7 TCP/IP Transmitter window

..........................................................................................................................................................

 